CT tube arcing is one of the most common reasons for sudden CT scanner downtime. If you hear a popping sound from the gantry or see an “Arcing” or “Tube High Voltage Error”, your tube is arcing. Here is what it means and what to do, from RIMION factory engineers.
What is CT Tube Arcing?
Arcing is an unwanted electrical discharge inside the X-ray tube insert. Instead of electrons flowing smoothly from cathode to anode, they create a spark through residual gas or contamination. This causes a high-voltage breakdown and the scanner stops exposure to protect itself.
5 Common Causes of Arcing
1. Gas Inside the Tube Insert Over time, metal parts release tiny amounts of gas. If vacuum level drops, gas ionizes and creates an arc. Common in old tubes or tubes stored for a long time without seasoning.
2. Cracked or Damaged Anode / Cathode Tungsten evaporation and thermal stress can create micro-cracks. Rough surfaces concentrate electric field and trigger arcing, especially at high kV (120-140 kV).
3. Contamination and Tungsten Deposition Tungsten from the filament deposits on the glass envelope. This contamination can cause small leakage paths and unstable high voltage.
4. Skipping Warm-Up and Seasoning Starting with high mA/kV on a cold tube is the fastest way to cause arcing. The tube needs gradual seasoning to re-absorb gas.
5. High Voltage Cable or Oil Problem Sometimes arcing is not in the insert but in the HV tank, cable, or oil. Contaminated oil or loose HV plug can create external arcing.
Symptoms of a Tube That is Arcing
- Loud pop or click sound from gantry
- Error: “Tube Arcing”, “HV Overcurrent”, “Tube 1 Overcurrent”
- Ring artifacts or white bands in images before failure
- Scanner needs frequent air calibrations
- Tube needs seasoning every morning
How to Fix and Prevent Arcing
Step 1: Run Tube Seasoning / Warm-Up Run the full manufacturer seasoning procedure for 10-15 minutes. Start low kV and increase gradually. This often removes light gas and stops arcing for weeks.
Step 2: Reduce kV and mA Temporarily If arcing happens at 140 kV, try 120 kV with lower mA. If arcing stops, the tube insert is aging and near end of life.
Step 3: Check HV Cables and Oil Inspect HV cable plugs for carbon tracks or oil leaks. Ask your engineer to check HV tank oil level and insulation. Many “arcing” errors are external.
Step 4: When to Replace the Tube If arcing happens more than 2-3 times per day, even after seasoning, the tube vacuum has failed. Continuing will damage the high-voltage generator. Time to replace.
Factory Tip from RIMION: Every RIMION tube is aged and tested at full HV before shipping to reduce early arcing.